Pular para o conteúdo principal
GE Healthcare · CTHiSpeed

Application software fails to boot up with insufficent disk space message

Symptom

Application software failed to boot with insufficent disk space message. Application software failed to boot with out of disk space message. See the long description for error codes from viewlog and SYSLOG Application software failed to boot with out of disk space message Thu Jun 23 07:02:15 2005 Error: 9071 Host: gtct3 Process: dbrserver File: proc.c Line: 376 get mail failed in process_mail() Thu Jun 23 07:02:15 2005 Error: 9056 Host: gtct3 Process: dbrserver File: server.c Line: 52 connection to Netserver failed, exiting... SYSLOG in /usr/adm: Jun 23 07:15:01 2C:gtct3 sendmail[2125]: NOQUEUE: SYSERR(root): queuename: Ca nnot create "qfHAA02125" in "/var/spool/mqueue" (euid=0): No space left on device Jun 23 08:28:45 3E:gtct3 ctuser: checkdisk: Disk partition /dev/usr is full (on gtct3).

Error message

watchswap: Disk partition /usr/g is full (on ct01). Error found in OC SYSLOG: root: watchswap: Disk partition /usr/g is full (on ct01). Found in cd /usr/adm more SYS*

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
102L.1_HSQXILINUXM3 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) (Model number: 2337535-2) - showprods returns 102L.1_HSQXILINUXM3 See LD for ClassC and ClassM part numbers 102L.1_HSQXILINUXM3 for HiSpeed QXi with Console (GOC2 / Linux Pegasus computer) (Model number: 2337535-2) Used on Product HiSpeed QXi with (Linux/Pegasus) Software Collector 2369632-2 CLASSC 2369625 CLASSM 2369626 showprods | grep Apps = 102L.1_HSQXILINUXM3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.