Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000740250444260007774260007937

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.

Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed.

Error: 260007937 DAS Converter Board(s) encountered a stage overflow during alpha calibration. See Long D

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 See LD for FEMC comments 2371627-4 replaces 2371627-3 GDAS / MDAS 16 Slice RIGHT Backplane Chassis Asm EDITION 1-2 Part #: 2371627-4 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS-MDAS 16 Slice RIGHT BP-Chassis Asm EDITION 1-2 FRU: Yes Status: Orderable Comments: Added as replacement for 2371627, -2, -3 & 2332630 per ecr 2012678, ms13452 050518 Part #: 2371627-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: BP-Chassis Right EMC Asm - GDAS FRU: Yes Status: Orderable Comments: 2371627-3 is backward compatible replacement for 2371627-2. DO NOT USE 2371627-2 to replace 2371627-3. 2371627-3 is EMC2 Compliant. ms13452 PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE MS8336 12/08/2004
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues