GE Healthcare · CTLightSpeed26000794026000777127259260007400109585207637776
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260007940 A calibration request failed to one or more converter boards See LD for complete error Error: 260007940 A calibration request failed to one or more converter boards Mon Apr 17 08:16:41 2006 Error: 260007940 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 500 Mon Apr 17 08:16:40 2006 Host : Das Ermes # : 260007940 Exception Class : Soft Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 500 Function : No System Function Reported Scan Type : None/Unknown Scan: 1094/2/0 A calibration request failed to one or more converter boards. Type : 3 Param 1: 6 Param 2: 1, a401 Error: 260007771 DAS Single Card Communication Error See LD for complete error Error: 260007771 DAS Single Card Communication Error (could be MDAS or GDAS) MDAS: Mon Apr 17 08:16:46 2006 Error: 260007771 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 863 Mon Apr 17 08:16:45 2006 Host : Das Ermes # : 260007771 Exception Class : Soft Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 863 Function : No System Function Reported Scan Type : None/Unknown Scan: 1094/2/0 M-DAS Single Card Communication Error. See Following Error Message for Converter Card List Board:19 Co
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2355552 MDAS 16 Convertor Card ( replaces 2322994-2 ) Used In LightSpeed 16 Gantry with MDAS. 2355552 MDAS 16 Converter Card used in LS 4X with 16 slice MDAS (replaces 2322994-2) Part #: 2355552 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.25 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SERVICE, MDAS16 CONV BD FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. This is an MDAS16 CONV BD .compared to the SDAS. Make sure the correct Convertor is Ordered and replaced.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required