Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260004025502291912726000104911497260001022

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260004025 DAS / Collimator firmware object: Lost communications with node applications. See Long Description for full error Thu Jan 13 06:58:23 2005 Error: 260004025 Host: OBC Process: Collimator File: col_cf_in_mail.c Line: 442 Function: No System Function Reported Scan: 50229/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:58:23:235 Log Series: 47 Collimator firmware object: Lost communications with node applications. Thu Dec 31 11:45:03 2009 Error:260004025 Host: OBC Process: DAS File: das cf in mail.c Line: 430 Function: No System Function Reported Scan: 19127/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:45:03:305 Log Series: 451 DAS firmware object: Lost communications with node applications.

Error: 260001049 Communications with the DCB sub-system has not been established. Scanning aborted. See LD for complete error Wed Nov 22 07:15:07 2006 Error: 260001049 Host: OBC Process: HOP File: hop_misc.c Line: 265 Function: No System Function Reported Scan: 11497/3/1 Type: Scout Exception Level: Secondary Time: 07:15:06:725 Log Series: 94 Communications with the DCB sub-system has not been established. Scanning aborted.

Error: 260001022 Co

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2289548 MDAS BACKPLANE/CHASSIS, CTR ASM used on 4/8 slice LS 3.x MDAS See long description for details Part #: 2289548 Part Type: GP CONSUMABLES Part Weight: 7.52 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS BACKPLANE/CHASSIS, CTR ASM FRU: Yes Status: Orderable Comments: FLS - MDAS CHASSIS, CENTER ASM-H3. USED ON WARP3. SUPPLIER IS CT MANF. ADD FOR ADMIN REASONS ONLY. WILL NOT BE SUPPORTED BY CA-GPUNTIL M3 (FW38-40). MSA6724 EJM 05/16/01 FOR CHASSIS REPLACEMENT, ORDER FLEX BOOT KIT PN 2329000THIS KIT CONTAINS 120 FLEX BOOT COVERS. ADDED PER CT SERVICE ENGINEERING (ERWIN SULMA) BY MS20544, CSV ON 1/20/02. SEE FE ASSIST. FLS - FE ASSIST: RETURN TO CT SERVICE ENGINEERING FOR EVALUATION. ATTN: TIM WATKINSGE MEDICAL SYSTEMS, 16800 W. RYERSON RD., NEW BERLIN, WI 53151. PH 262 785-4221.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.