Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1831212257550013

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183121 FIL/MA fault detected: Open filament circuit. See the long descripton for the full error code Tue Oct 11 13:12:27 2005 Error: 183121 Host: OBC Process: Generator Function: XRAY GENERATION : Filament Selection Scan: 22575/2/1 Type: Helical Exception Level: Pri/Most Severe FIL/MA fault detected: Open filament circuit. Desired focal spot = 0. (0 = large, 1 = small) <== Large Filament is open OR Fri Sep 22 07:03:11 2006 Error: 183121 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 341 Function: XRAY GENERATION : Filament Selection Scan: 50013/10/40 Type: Axial Exception Level: Pri/Most Severe Time: 07:03:11:045 Log Series: 393 FIL/MA fault detected: Open filament circuit. Desired focal spot = 1. (0 = large, 1 = small)<-------- Small

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-136370P23 K78 FILAMENT RELAY See long description for details Part #: 46-136370P23 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 024VDC 2P2T 028VDC 10AMP FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues