GE Healthcare · CTLightSpeed260007391
Scanning aborted with DAS power errors
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391The Data Acquisition System Has Detected An Error In The +5V DCB/DAS Digital DAS Power Supply. Wed Apr 5 10:58:29 2006 Error: 260007391 Host: orp Process: orp_subsys File: pwrtst.c Line: 90 Wed Apr 5 10:58:28 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The Data Acquisition System Has Detected An Error In The +5V DCB/DAS Digital DAS Power Supply.^B Measured: 0 mV, Expected range: 4750 to 5250 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 See long description for details 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 Part #: 5126615 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Digital PS Assembly FRU: Yes Status: Orderable Comments: Added per RT / GDAS RSPL
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.