GE Healthcare · CTLightSpeed2601000962300000675000416800015000
Scan hardware faults
Symptom
Intermittent Scan Hardware fault
Error message
Error: 260100096 Exposure Interlock Opened See Long Description for complete error. 260100096 Exposure Interlock Opened Mon Dec 6 09:54:17 2004 Error: 260100096 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 546 Mon Dec 6 09:54:17 2004 Host : Table Gantry Processor Ermes # : 260100096 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 546 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Axial Scan: 7629/2/1 Exposure Interlock Opened. The current loop from Gantry to PDU. The cause of the failure: Console Relay Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx L
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 ft. 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 ft. See LD for picture 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 f t. Part #: 2367805 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OUTLET BOX ASSY - GOC/GRE FRU: Yes Status: Orderable Comments: FLS - CABLE BULKHEAD ASSY (GRE) USED ON H16 ADDED PER GRE RSPL MS13452 RISK: FUNCITONALITY FREQUENCY: LOW Mark Number FRU Description 4 2367805 Yes CABLE BULKHEAD ASSY (GRE)
- Area
- CT
Full cause & step-by-step resolution