GE Healthcare · CTLightSpeed18280450080

Gantry Display blank

Symptom

Gantry Display blank. Gantry Display blank

Error message

Error: 182804 The Gantry Display is NOT Initialized Properly See the LD for the full error code Error: 182804 The Gantry Display is NOT Initialized Properly Wed Feb 8 07:20:05 2006 Error: 182804 Host: ETC Process: Scan File: etc_in_mail.c 1.38 4/16/01 Line: 648 Function: No System Function Reported Scan: 50080/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 07:20:05:645 Log Series: 37 HW/SW error: Communication Problems with the Gantry Display. The Gantry Display is NOT Initialized Properly. (Scanning is NOT allowed if the Gantry Display is NOT Working)

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.