GE Healthcare · MROvation2224709
Signal Drop off or loss on CTL Coil Due to defective Mousehouse.Replace Mousehouse.
Symptom
Intermittent Signal Drop off or loss on the CTL Coil The problem happens with HFO3 M3 and M4 software. Only the Thoracic and Lumbar are affected. The affected images look noisy.
Error message
2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver (X + 1) while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to occur.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- MFO2 Ovation2 with Octane2
- Component
- 2218400-2 MOUSE HOUSE (MFO)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.