Defective IT Fiber Optic cable between the Host computer and the System Cabinet.

Symptom

Can Not Scan

Error message

2252055 IT-Host Card driver initialization error. This error usually means that the IT cable is disconnected or that the cable is damaged

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
Component
2318811 IT-Link Cable Run # 1044 OW1 A15 BIT3 TO MR2 A11 J22
Area
MR