Pular para o conteúdo principal
GE Healthcare · CT200002352230011004200002368

Lightspeed 1.X - Octane

Symptom

System will not boot to or stay in applications. System will not boot or stay in applications

Error message

Error: 200002352 A HOST_SHUTDOWN was initiated by the death of scanmgr See Long Description for full error Tue Nov 23 17:26:47 2004 Error: 200002352 Host: srct Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiated by the death of scanmgr With 06MW03.X software this is listed as: Fri Oct 6 19:02:31 2006 Error: 200002352 Host: dominct Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanmgr.

Error: 230011004 Unable to initialize scan data disk handling. The size of the scan data disks appears incorrect See the LD for the complete error Error: 230011004 Unable to initialize scan data disk handling. The size of the scan data disks appears incorrect. Mon Apr 2 08:00:48 2007 Error: 230011004 Host: ct01 sbc Process: scanfilemgr File: sdio xfs api.c Line: 2351 Function: System Control : Scan Database Manager Scan Type: Not scanning Scan: 0/0/0 Unable to initialize scan data disk handling. The size of the scan data disks appears incorrect. Number of extents: 0 Number of blocks: 0 Geometry: 200h:200h:80h This can occur if there is no power to the scan data disks.

Error: 200002368 A set of processes ar

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.