Unable to Scan with Multi Coil TR Driver Faults.

Symptom

Unable to Scan with Multi Coil TR Driver Faults

Error message

2225552 Multi-Coil T/R Driver Board Fault. An Open Circuit was Detected on Port 2. Thu Aug 21 12:06:10 1997 Suite: MRS1 Host: OC0 Proc: HPC_PORT_ID Error: 2225552 File: PMH.m 1.29 Method: spi_tr_mon Line: 6240 Multi-Coil T/R Driver Board Fault An open circuit was detected on port 2. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.