GE Healthcare · MROvation2225476
Defective Analogic RF amplifier.
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2225476 RF Amplifier failed to power off in the allotted amount of time when commanded to do so.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Component
- 2320323 0.35T TELI RF AMP 2320323 0.35T TELI RF AMP 2320323 0.35T TELI RF AMP
- Area
- MR