Infinia MWL acquired study, the date and or time shows up wrong when transferd to PACS Study Time (0008,0030)…

Infinia MWL acquired study, the date and or time shows up wrong when transferd to PACS

Symptom

Infinia MWL acquired study, the date and or time shows up wrong when transferd to PACS Study Time (0008,0030) and Study Date (0008,0020) DICOM tags are not updated properly when Infinia uses modality worklist. The values of Study Date and Study Time get set to the values of the Scheduled Study Start Date (0032,1000) and Scheduled Study Start Time (0032,1001) tags received from the modality worklist provider. The Infinia software does not update Study Date and Study Time when the first acquisition takes place.

Applies to

Models
Infinia, Xeleris
Product
Infinia II Infinia II Acquisition System
Software
Infinia All Infinia All. Including Infina 1 and Infinia II. Infinia All. Including Infina 1 and Infinia II.
Component
*Software Software Software
Area
Nuclear Medicine

Full cause & step-by-step resolution