GE Healthcare · CT26000240339667260002318
Scanner hardware fault
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260002403 The Data Acquisition System Has Detected An Error In The -5V or +5V Analog DAS Power Supply. See Long Description for the full error code Tue Dec 7 07:20:51 2004 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 783 Function: DATA ACQUSITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:20:50:240 Log Series: 23 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply. Measured: 508 mV, Expected range: -5250 to -4750 mV Fri Jul 17 16:00:53 2009 Error: 260002403 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 729 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 39667/9/1 Type: Helical Exception Level: Sec/Soft Time: 16:00:54:485 Log Series: 357 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply.^B Measured: 4732 mV , Expected range: 4750 to 5250 mV Error: 260002318 A Failure Has Been Detected By A DCB Selftest. DCB Power Supply selftest failed See the Long Description for the full error code Error: 260002318 A Failure Has Been Detected By A DCB Selftest. DCB Power Supply selftest failed Tue Dec 7 07:20:51 2004 Error: 260002318 Ho
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218158 SDAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET. used on 1.x with H1 gantry See Long Description for details Part #: 2218158 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS 5VDC ANALOG PWR SUPPLY W/BRACKET FRU: Yes Status: Orderable Comments: FLS - DAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET U/O HELIOS DAS REPLACES 46-251198P46 MS08992 10/5/98 GENCO INDUSTRIES-WAUKESHA,WI PER FE - 46-251198P46 IS NOT ENTIRELY COMPATIBLE - 3 CONNECTION POINTS FOR WIRES - NEED 4- CANNOT SOLDER WIRE IN CORRECTLY DUE TO HOLE IN SOLDER POINT BEING TOO SMALL FOR MORE THAN 1 WIRE. TW/01/03
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required