Pular para o conteúdo principal
GE Healthcare · MROvation

Software was corrupted due to a bad block on the hard drive.

Symptom

HP Computer Fails to Boot Linux Operating System.

Error message

I/O error in file system SDB2 Metadata device SDB2, Block xxxxxx I/O error in file system ("SDB2"). Metadata device SDB2, Block xxxxxx (where xxxxx is the block number). This error appeared in the boot-up messages which scroll during software boot. Boot process halts at this message. Does not open a separate window.

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OVATION5 Applies to Ovation5 (Excite) only. Ovation5 is 12.x platform software, not OvationHD.
Component
5143798-5 250GB SATA 150 7200 RPM HDD
Area
MR