Scans Intermittently Stop With Coil ID Errors, Coil ID LED turns Red

Symptom

Intermittent Coil ID Errors in Middle of Scans Coil ID LED turns Red This problem began right after the MSHS was replaced for a different problem. The system recognizes the coil and allows the scan to begin but then the system intermittently stops the scan and logs Coil ID errors, and the Coil ID LED turns red. This happens with all coils. The customer has to shut the system down and power cycle the whole system to fix the problem. Then later this process repeats. This problem began right after the MSHS was replaced for a different problem. The system recognizes the coil and allows the scan to begin but then the system intermittently stops the scan and logs Coil ID errors, and the Coil ID LED turns red. This happens with all coils. The customer has to shut the system down and power cycle the whole system to fix the problem. Then later this process repeats.

Error message

2219986 Coil ID signal received while Scanning, Coil Code : [0] 2219986 BROCKGEMR44 scn SrxCoilID.m 455 Coil ID signal received while Scanning, Coil Code : [0]

2218624 Failed to execute the next event. Error: 2218624 Host: lx-mr Process: scn File: StateMachine.c Line: 200 Failed to execute the next event.

Applies to

Models
Ovation
Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
2218400 HFO3 MSHS Mouse House T3503RT
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required