GE Healthcare · MROvation22333112233312
CERD is Failing to Download to the CAP.
Symptom
UCERD is Failing to Download to the CAP. Internal network errors occur intermittently when downloading to the CAP on the UCERD Digital Board. A workaround is to cycle power to the TPS and download it to restore communication.
Error message
2233311 CERD NETWORK ERROR-CAP Ignored CERD NETWORK ERROR\n\ The TPS has placed a packet in DPR for the CAP to take, but the CAP has completely ignored it. The CERD DPR may be broke or (most likely) the network software is not running on the CAP. Reset and/or power-cycle the TPS chassis. 2233312 CERD NETWORK ERROR-CAP Responded The CAP has responded to a network packet transaction with the TPS but the response is invalid. The network software on the CAP is probably not running. Reset and/or power-cycle the TPS chassis.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.