CERD is Failing to Download to the CAP.

Symptom

UCERD is Failing to Download to the CAP. Internal network errors occur intermittently when downloading to the CAP on the UCERD Digital Board. A workaround is to cycle power to the TPS and download it to restore communication.

Error message

2233311 CERD NETWORK ERROR-CAP Ignored CERD NETWORK ERROR\n\ The TPS has placed a packet in DPR for the CAP to take, but the CAP has completely ignored it. The CERD DPR may be broke or (most likely) the network software is not running on the CAP. Reset and/or power-cycle the TPS chassis.

2233312 CERD NETWORK ERROR-CAP Responded The CAP has responded to a network packet transaction with the TPS but the response is invalid. The network software on the CAP is probably not running. Reset and/or power-cycle the TPS chassis.

Applies to

Models
Ovation
Product
Ovation/MFO
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.