Symptom

Scanner computer hangs intermittently

Error message

Error: 260005022 procICE may have received a late ping response from ICE. See LD for complete error Wed Dec 28 06:56:50 2005 Error: 260005022 Host: rct Process: procICE File: iceReceive.c Line: 273 procICE may have received a late ping response from ICE. Last ping message from procICE information: sent time : Wed Dec 28 06:55:46 2005

Error: 260005003 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Thu Apr 29 06:05:11 2004 Error: 260005003 Host: fitzipct Process: procICE File: iceCpu.cxx Line: 343 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Last ping was at Sun May 30 22:21:46 2032 with seq num 1818850152 iceConsole.log: Initializing System Memory (DRAM)... System Memory: 128MB, ECC Enabled (ECC-Memory Detected) PPC1-Bug

Error: 182119 LAN Watchdog TIMEOUT for the SBC. See Long Description for more details Sun Sep 17 10:24:42 2006 Error: 182119 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 556 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe 

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2174128 Ethernet cable. The description calls the cable the Ethernet switch to Transceiver cable, but it is just a general purpose Cat5 Cable that is used in many places. See LD for part information Part #: 2174128 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: HELIOS ETHERNET SWITCH TO FRU: Yes Status: Orderable Parts Offered as: Current: 2174128 Harvested: Not Available Exchange: Not Available Comments: FLS - HELIOS ETHERNET SWITCH TO TRANSCEIVER U/O HELIOS-CONSOLETRANSCEIVER 10/8/98 MS08992 M3 OCT '98 CABLE MATERIAL CHANGED FROM QUABBIN 9688 TO QUABBIN 2908 PER PCN 220402. USE UP EXISTING STOCK. ADDED BY MS20544, CSV ON 1/10/02.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.