Low signal on Head coil. 1000V bias voltage not present.
Symptom
Low signal on Head coil. 1000V bias voltage not present. Unable to run head calibrations. Running software MFO OPENMR4.292a_M4_0433.f. Site could not get much signal from the head coil, all others ok. Could not proceed to calibrate the system. The signal was being coupled away by XMIT coil. T/R switching signals OK and 1000 volts was available but not for head scanning.
Error message
2252232 SCP detected a Node Reset on the CAN network Reset Node is Drivers Control Board 1093421090 2 1 Wed Aug 25 08:04:50 2004 2252232 2 MR-OC0 NSP SCP:CanLinkHWO 595 SCP detected a Node Reset on the CAN network. Reset Node is Drivers Control Board. EN 183 SR 209
Applies to
- Models
- Ovation
- Product
- Ovation with EXCITE
- Software
- OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required