Clarify #033367 No Signal With Body Flex Coils

Symptom

An HFO3 system could not get any signal from either the Body Flex L or Body Flex M coils. When the customer or FE enters Manual Prescan with these coils plugged in, the power spectrum is just a flat line. All other coils are working fine. The FE has replaced the two Body Flex coils and the MSHS module without fixing the problem.

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

Applies to

Models
Ovation
Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
2218400 HFO3 MSHS Mouse House T3503RT
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.