Clarify 031483 OVATION Important Note About the RF Switch Box and Body Hybrid
Symptom
Damage Will Result to the RF Switch Box and Body Hybrid. Damage will result to the RF Switch Box and Body Hybrid if the high voltage is turned on at the SSM and if any one of the RF (black cable) xmit/receive cables is disconnected either at the body coil or at the RF Switch Box on the Ovation.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- OVATION5 Applies to Ovation5 (Excite) only. Ovation5 is 12.x platform software, not OvationHD.
- Area
- MR