Pular para o conteúdo principal
GE Healthcare · CT230019000230011034

System will not stay up in applications and shutsdown

Symptom

System will not stay up in applications and shutsdown. System will not stay up in applications and shutsdown

Error message

Error: 230019000 The initialization of the scan disk I/O proxy layer failed.

Error: 230011034 Unable to initialize scan data disk handling Thu Sep 4 12:04:35 2003 Error: 230011034 Host: imcct Process: scanfilemgr File: sdio_xfs_io_ops.c Line: 1228 Function: System Control : Scan Database Manager Scan Type: Not scanning Scan: 0/0/0 Unable to initialize scan data disk handling. Unable to create information file. Info file: /raw_data/info_file Mode: c2h Permissions: 667 (octal) Unix system error : 9 --- Bad file number.

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2191454-3 SCSI Cable used on LightSpeed H1& H2 Systems See LD for details Part #: 2191454-3 Part Type: GP CONSUMABLES Part Weight: 0.91 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI CABLE, DRV BOX TO S/BLKHD, HD68 FRU: Yes Status: Orderable Comments: FLS - SCSI CABLE, DIV BOX TO S/ BH HD68. REPLACES 2191454-2 PER PCN 199904. USE UP OLD FIRST.USED ON H1 AND H2. MS26561 DSM 11/7/00 *PER CT MFG - THEY NO LONGER STOCK THIS AT CT AS IT IS OBSOLETE TO CT - 1/29/03 - MS22763-JP* MUST OBTAIN FROM NORTECH LightSpeed QXi with no O2 or original H2 Systems with HSD in Dog cage: Bulkhead J51 to Drive Box (Molded) is 2191454-3
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues