GE Healthcare · CT245071102900024555158001182122
Scanner Hardware Stopped Scan
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 245071 The System Aborted Scanning:, views expected 984 - views collected 0 See the LD for the full error code Tue Jan 3 16:54:34 2006 Error: 245071 Suite: LCT1 Host: SBC0 Process: scanmgr File: do_axial_complete.m 1.16 Method: do_axial_complete() Line: 582 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Xray-on Axial Scan Seq: 1029000 Axial Scan Status Message: The System Aborted Scanning:, views expected 984 - views collected 0 Error: 245551 FEP requested scan abort See the long description for the full error code Tue Jan 3 16:53:23 2006 Suite: LCT1 Host: SBC0 Proc: dataacq Error: 245551 File: tl_fep_events.m 1.18 Method: event_fep_abort_xr Line: 260 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Xray-off Axial RTS Scan Seq: 58001 FEP requested scan abort Error: 182122 Received RP/FEP xray abort request from uuid 304 See the LD for the full error code Tue Jan 3 16:48:37 2006 Suite: LCT1 Host: STC Proc: Master Exc Handler Error: 182122 File: exm_in_mail.c Method: No Method Line: 729 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 16:48:36:440 Log Series: 130 INFO: Received RP/FEP xray
Applies to
- Product
- HSA RP 1.X HSA RP 1.X with Zeus HSA Gantry and Sun Genesis Console FMI 25246 HSA/RP 1.6 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-183027P90 E-DAS TIMING CARD See long description for details Part #: 46-183027P90 Part Type: GP REPAIRABLES Part Weight: 0.45 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: E-DAS TIMING CARD FRU: Yes Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.