Pular para o conteúdo principal
GE Healthcare · CT200920002593850972150230017112230017810800000

Lightspeed 1.X - Octane

Symptom

Unable to calibrate system after performing fast tube change

Error message

Error: 200920002 hhs tool determined the aperture specs are not met test result inf specs See LD for full Error Thu Dec 1 16:29:27 2005 Error: 200920002 Host: gect Process: hhs File: hhsUtils.c Line: 300 hhs tool determined the aperture specs are not met test result inf specs [3.593850 - 3.972150]

Error: 230017112 Accs server, hhs_dose processing failed the sanity check of scan data See LD for Full Error Thu Dec 1 16:29:34 2005 Error: 230017112 Host: gect_sbc Process: accs File: do_hhs.c Line: 144 Accs server, hhs_dose processing failed the sanity check of scan data

Error: 230017810 ACCS CCal: Error: Outer Data Row may be in penumbra See LD for Full Error Thu Dec 1 16:13:04 2005 Error: 230017810 Host: gect_sbc Process: accs File: collSweepProcessing.c Line: 247 ACCS CCal: Error: Outer Data Row may be in penumbra. Check for excessive BOW alignment error or increase cam range during calibration. Limit 0.800000, Computed 0.353240

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues