GE Healthcare · CT230017801230017812000500
Scanning aborted with collimator errors
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1 Error: 230017812 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500 See Long Desciption for complete error Mon May 30 05:41:49 2005 Error: 230017812 Host: cxi1 Process: accs File: collSweepProcessing.c Line: 491 ACCS CCal: Error: too much variation in ratio data. Limit 0.000500
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required