GE Healthcare · CTDiscovery260107413
Scanner Hardware Stopped Scan.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260107413 The Axial AC Drive reported fault code: F:13 Ground Fault See long description for full error Thu Jul 24 09:08:06 2008 Error: 260107413 Host: tgp Process: tgp_subsys File: AX2fault.c Line: 195 Thu Jul 24 09:08:06 2008 Host : Table Gantry Processor Ermes # : 260107413 Exception Class : Abort Severity : Primary File : AX2fault.c Line# : 195 Function : AXIAL CONTROL : Axial Communications Scan Type : Scout Scan: 193/1/1 %s The Axial AC Drive reported fault code: F:13 Ground Fault Check the motor and external wiring to the drive output terminals for a grounded condition.
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2233402 Axial Drive Motor Controller See Long Description for details Part #: 2233402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 8.64 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AX DRIVE MTR CONTRLLER W/CORERS ASSY FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONICS. MS26561 DSM 1/7/01 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. DO NOT DISCOUNT - PER SHAKUN*** 2ND OPINION NOTES *** Check fuses/connections associated with axial drive circuit; run axial functional diagnostic test and check for failures. If used on HP60
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required