All three XPSs show cold-plate over temps due to check valve installed backwards.

Symptom

Cannot scan with XPS cold plate over temperatures on all three axis

Error message

2267841 The % axis XPS is reporting a cold plate over-temperature The cold-plate temperature is 56C .

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.