GE Healthcare · MROptima2269060198232267728226741425411
XGD Not Found Due to 48 Volt Power Supply(s)
Symptom
Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.
Error message
2269060 The 1-Wire Network did not detect XGD GMC Power. Error: 2269060Host: PHSRADMR02 Process: diagCoord File: oneWire.c Line: 1143 Task ID: 0x6db4420 Time: 19823 Object: AGPException Class: NotificationFunction: Software/Firmware/Config: N/AThe 1-Wire Network did not detect XGD GMC Power 2267728 Unexpected communication loss with XGD. Please TPS reset and try again. If this error occurs again, contact your service representative. See Long Description ERROR: EM_XGD_HWO_COM_FAILURE Extended Error: The SCP unexpectedly lost its Ethernet connection to the XGD. Verify that an Ethernet cable is connected from the GMC to the switch and that the link LED on the GMC Ethernet port is illuminated (top LED). Verify that an Ethernet cable is connected from the SCP to the switch and that the link LED on the SCP Ethernet port is illuminated (top LED). Power down the XGD and verify that the GMC is seated securely in its slot by removing and re-seating it. Power on the XGD and perform a TPS reset. Power down the MGD chassis and verify that SCP is seated securely in its slot by removing and re-seating it. Power on the MGD chassis and perform a TPS reset. 2267414 NO XGD MODULE FOUND. The IRF3 bo
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Component
- 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.