Pular para o conteúdo principal
GE Healthcare · MROptimaDiscovery2261715

Cannot Scan with Gradient Coil Coolant Resistivity Alarms.

Symptom

Cannot Scan with Gradient Coil Coolant Resistivity Alarms Cannot Scan with Gradient Coil Coolant Resistivity Alarms.

Error message

2261715 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. See Long Description Error: 2261715 Host: lx-mr Process: NSP File: SCP:CoolingHWO Line: 2901 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FSE-XL Coil: HDCTL 456 Scan: 521/9 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. Scanning will be inhibited until coolant serviced. Please contact service.--> Extended Error: Order coolant deionization kit and schedule service. If the resistivity is 0 kOhm-cm, check the wiring between the sensor in the gradient coolant reservoir and the signal box.

Applies to

Models
Optima, Discovery
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required