GE Healthcare · MROptimaDiscovery2261715
Cannot Scan with Gradient Coil Coolant Resistivity Alarms.
Symptom
Cannot Scan with Gradient Coil Coolant Resistivity Alarms Cannot Scan with Gradient Coil Coolant Resistivity Alarms.
Error message
2261715 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. See Long Description Error: 2261715 Host: lx-mr Process: NSP File: SCP:CoolingHWO Line: 2901 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FSE-XL Coil: HDCTL 456 Scan: 521/9 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached an alarm level. Scanning will be inhibited until coolant serviced. Please contact service.--> Extended Error: Order coolant deionization kit and schedule service. If the resistivity is 0 kOhm-cm, check the wiring between the sensor in the gradient coolant reservoir and the signal box.
Applies to
- Models
- Optima, Discovery
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.