Gradient error 2267857 & 2267827 due to XGD amplifier.Scan Pauses with Gradient Errors and Continues after start Scan Pr…
Gradient error 2267857 & 2267827 due to XGD amplifier.Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact.Replace XGD Amplifier wqr.Optima MR4…
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2267827 XGD hardware warning detected. The %s bit in %s register was detected on the %b %b. See Long Description Error: 2267827 Host: SLMAINMR Process: NSPFile: SCP:FspMac.cpp Line: 320Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Gradient Echo Coil: 8HRBRAIN Scan: 543/2 XGD hardware warning detected. The ovrSR bit in DSP_FAULT register was detected on the XGD XGA-CB board % axis.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Component
- 5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.