GE Healthcare · MROptima22678922267885
2267885 The Z axis XPS is reporting a PLL Warning Due to XGD Power Supply.Replace XGD Power Supply XPS 5148804-2.
Symptom
Will Not Scan with Gradient Errors.
Error message
2267892 The % axis XPS is reporting an Int4_Lock Fault. This is normal during a TPS Reset, but should not occur during scan-time See Long Description 2267892 MOI_MR NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: HNS CTL456 Scan: 129/7 The % axis XPS is reporting an Int4_Lock Fault. This is normal during a TPS Reset, but should not occur during scan-time.-> Extended Error: Ensure that J4 clock fiber is connected to the XPS. Ensure that GMC has the MR System Clock plugged into J7. Ensure that the optical output power of the CB If still failing only on this fault, it indicates a PLL failure on the XPS-SB. 2267885 The % axis XPS is reporting a PLL Warning See Long Description Error: 2267885 Host: signa2 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The % axis XPS is reporting a PLL Warning.-> Extended Error: Ensure the fiber optic cable J4 is connected to the corresponding XPS, and that light is present. Check that the corresponding channel on the XPS-CB in the Cronus chassis has a good fiber optic output.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.