GE Healthcare · MROvation20000236822238222223817
Fails to Boot Applications and TPS Download Fails Due to TPS/ISE Backplane Power Supply.
Symptom
System fails to boot Applications software and TPS reset fails. See long description. The system is a 1.5T LX, running 9.1 M4 software. The system will not boot applications. Looks like the Bit3 is down. Errors in the log look like TPS reset failure. Ping to tps fails as seen below: lx-mr 6# ping lx-mr-tps PING lx-mr-tps (216.33.27.2): 56 data bytes ping: sendto: Connection refused
Error message
200002368 A set of processes are deadlocked waiting for each other to start. Mon Apr 12 11:43:42 2004 Error: 200002368< Host: BaptistGEMR Process: hostMonitorbsp File: ssProcStart.c Line: 1099 A set of processes are deadlocked waiting for each other to start. svat, scn, rtcProxy, resetServer, lais, hgp, displayServer, autoview, app, WLServer, TIR, SITT, RtpHost, RTPSIServer, RIR, PPSproxy are the likely culprits. This deadlock has caused the system to shutdown! 2223822 Reset TPS was not successful. Check System Cabinet power and connections via BIT3 interface. Sat Nov 6 07:30:31 2004 Error: 2223822 Host: lx-mr1c Process: NSP File: TardisReset.cc Line: 522 Reset TPS was not successful. Check System Cabinet power and connections via BIT3 interface. 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.
Applies to
- Models
- Ovation
- Product
- Ovation/MFO
- Software
- MFO2 Ovation2 with Octane2
- Component
- 2118408-3 MAIN POWER SUPPLY - BACKPLANE This is the TPS/ISE Backplane Power Supply. It is referred to in Renewal Parts as MAIN POWER SUPPLY – BACKPLANE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100060 Unable To Film.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.
- Having Difficulties Setting Up Gradcal Scan From Dir 5112585 - Signa Ovation With Excite Technology
- CWA ID101091. Some or All Images in the Last Scan Did Not Reconstruct.
- HFO Cradle Will Not Move Due To Loss Of Home Flag Position.