Auto and Manual Prescan Fails.

Symptom

Auto and Manual Prescan Fails

Error message

2249173 Multi-Coil T/R Driver Fault. An open circuit was detected on port 2. 2249173 Multi-Coil T/R Driver Fault An open circuit was detected on port 2. Check cable to multi-coil board or check for an open pin diode.

Applies to

Models
Ovation
Product
Ovation with EXCITE
Software
OVATION5 Applies to Ovation5 (Excite) only. Ovation5 is 12.x platform software, not OvationHD.
Component
2160200-45 SSM assy for Ovation/OpenSpeed
Area
MR

Full cause & step-by-step resolution