The Center Frequency Was Out of Specified Range when Set Manually in Manual Prescan and Saved.
Symptom
Auto and Manual Prescan Fails
Error message
2267430 Bad Center Frequency Requested. The DTX-0 module was commanded to a center freq of XXXXXXXX Hz 2267430 Host: MR450 Process: diagCoord File: IrfRmtDev.cpp Line: 824 Task ID: 0x4cadc30 Time: 140334 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A Bad Center Frequency Requested.The DTX-0 module was commanded to a center freq of 6385980 Hz. The module is limited to 61863850 to 65863850 Hz. This is a software error.->
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR