Error Code 2261714 is Being Logged Periodically due to the HEC Signal Box.
Symptom
2261714 is Being Logged Periodically. This is Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. See Long Description The error message shown below is being logged periodically. It's being logged every few minutes. What does this mean and how do we get it to stop being logged? Mon Sep 15 05:55:48 2008 2261714 LX-BAY11B NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning->
Error message
2261714 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning See Long Description Mon Sep 15 05:55:48 2008 2261714 LX-BAY11B NSP SCP:CoolingHWO 2957 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant resistivity has reached a warning level. Please contact service. This alarm will not stop scanning-> Extended Error: Order coolant deionization kit and schedule service. If the resistivity is 0 kOhm-cm, check the wiring between the sensor in the gradient coolant reservoir and the signal box.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5269691 Signal Box kit Signal Box kit 5269691
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.