Symptom
No images being displayed after scanning a patient. No images being displayed after scanning a patient
Error message
Unrecovered data block read error in the /usr/adm/SYSLOG Unrecover data block read error in the /usr/adm/SYSLOG Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: retrying request Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: [Alert] Media error : Unrecovered data block read error (asc=0x11, asq=0x0) , (data byte 36) , Block # 1102732 (7041932) Unrecovered data block read error in the /usr/adm/SYSLOG Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: <6> retrying request Nov 29 12:25:21 6A:anact03801 unix: dksc0d1: <6>[Alert] Media error<6>: Unrecovered data block read error (asc=0x11, asq=0x0)<6>, (data byte 36)<6>, Block #1102732<6> (7041932)<6> Error: 230006528 ReconMgr sent abort See LD for full error Sat Mar 12 20:07:39 2005 Error: 230006528 Host: ice Process: reconmgr File: rm_process_job.c Line: 844 Function: Image Generation : SBC Processing Job Type: Scout Scan: 30771/1/1 Image: 30771/1/1 Recon Seq Id: 28 ReconMgr sent abort (recon_seq_id=0) to IC, got an unsuccessful status=-1. Error: 230006516 Reconmgr failed twice to send abort to IGCtrl, SdcCtrl and IC. Ignored See LD for full error Sat Mar 12 20:07:39 2005 Error: 230006516 Host: ice Process: reconmgr File: rm_pipeline.c Line: 28
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-78UU or 2169940-78 Octane/Octane II Computer 18 GB Disk Drive. See long description Part #: 2169940-78UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Not Available Status: Orderable Stan Pearson, Asset Mgmt move to used material program 6-27-05 Part #: 2169940-78 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Yes Status: Orderable Comments ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 0711261b. (212001261 8/20/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.