GE Healthcare · Nuclear MedicineDiscovery
What is the criteria to determine CZT detector "energy threshold" for bad pixels?
Discovery NM 530c/570c
Symptom
What is the criteria to determine CZT detector "energy threshold" for bad pixels? ' Energy threshold ' is new CZT pixel failure mode introduced with software version 1.003.150.3
Applies to
- Models
- Discovery
- Product
- Discovery NM 530c/570c Discovery NM 530c/570c (CZT) DSV NM 530c DSV NM 570c
- Component
- *CZT Image Detector *CZT Image Detector
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- VG Fanbeam collimators copying cor correction values
- Hawkeye X-Ray patient images ring artifacts
- Ventri failed system startup test, "LAST AXIS MOVEMENT FAILED"
- VG detectors radial 1 or radial 2 will not radial out in fast mode with RCU
- Ventri CAN Network Maping Initialization - ERROR_GCU_CAN_NETWORK_INITIALIZATION
- Venti or Discovery 530/570 Internal Host, DIB and GCU IP Address