HEC Coolant Flow Low Due to the return valve shut.

Symptom

Errors 2261712 and 2261723 HEC Coolant Flow Low are Logged and Scanning Not Inhibited See Long Description Errors 2261712 and/or 2261723 HEC Coolant Flow Low are being logged and scanning is not inhibited. One or both of these errors appear repeatedly in error log but scanning not inhibited: Wed Jun 30 08:49:36 2010 Error: 2261712 Host: XXXXXXXX Process: NSP File: SCP:CoolingHWO Line: 2942 Exception Class: Unknown Severity: Unknown Function: Power/Support Systems: Heat Exchanger PSD: Gradient Echo Coil: HEAD Scan: 2/2 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant flow is too low. Verify that the pump is turned on, all in-line valves are open. This alarm will stop scanning-> Wed Jun 30 08:49:36 2010 Error: 2261723 Host: XXXXXXXX Process: NSP File: SCP:CoolingHWO Line: 2942 Exception Class: Unknown Severity: Unknown Function: Power/Support Systems: Heat Exchanger PSD: Gradient Echo Coil: HEAD Scan: 2/2 Heat Exchanger Cabinet (HEC) detected the power electronics coolant flow is too low. Verify that the power electronics pump is turned on, all in-line valves are open. This alarm will stop scanning->

Error message

2261723 Heat Exchanger Cabinet (HEC) detected the power electronics coolant flow is too low. See Long Description Verify that the power electronics pump is turned on, all in-line valves are open. This alarm will stop scanning-

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.