Pular para o conteúdo principal
GE Healthcare · MRDiscovery22730002273010

Coil validation error due to bent pins in the coil connector.

Symptom

Coil ID Fails and CoilID Light is Red See Long Description Customer is attempting to scan, but getting error that something is wrong with coil ID. The Coil ID LED turns red.

Error message

2273000 Coil ID device was not detected on a coil see long Error: 2273000 Host: TICMR2 Process: NSP File: SCP:EclTask Line: 799 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Connector: Port 1 (P1

2273010 Coil interface detects a short on one-wire bus Coil communication error: Coil interface detects a short on one-wire bus. Check coil connection. Connector: %b Extended Error: This could indicate a bad connection at the coil interface, a bad coil, bad cables to the RF Hub RFCB, or a bad RFCB. Try connecting another coil. Run the Coil ID Diagnostic or Coil Datapath Diagnostic to troubleshoot cabling issues and to isolate the coil interface from the coil.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5339708-3 3T 32Ch Torso Coil - Posterior with cable
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.