GE Healthcare · MROptimaBrivo2256854

ICN VRE configuration fails time out at waiting 360 seconds for NFS mount request due to defective ICN.

Symptom

TPS reset fails with 2256854 errors Process DPx on board ICNx appears to have failed

Error message

2256854 Server Name: DS Process DP on board ICN1 appears to have failed. Please TPS reset to recover. After LFC and ICN Reconfigure, error persists

Applies to

Models
Optima, Brivo
Product
Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
Software
SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
Component
5337894-10 ICN X4170 used on Optima 360 and Brivo 355
Area
MR