Pular para o conteúdo principal
GE Healthcare · MROptima226783452532

Safemode Startup caused by a defective Z axis fan tray.

Symptom

Unable to Scan with Safemode Startup of XGD Failed for each Axis See Long Description Unable to scan with errors logged "Safemode Startup of XGD Failed" for every gradient axis seen at same time in error log. See errors, below. Notice that the date/time is the same for each error, and that all axes are reporting the error: X, Y & Z. Fri Mar 27 11:31:06 2009 Error: 2267798 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267801 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267804 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267807 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267810 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267813 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .->

Error message

2267834 The Power Cabinet AUX board is reporting a fan tachometer failure on # axis See Long Description Error: 2267834 Host: B4 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: mcgrafidy Coil: MCGRAFIDY1-6 Scan: 52532/1 The Power Cabinet AUX board is reporting a fan tachometer failure on # axis. Where # is X, Y, or Z axis.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5183212 FAN TRAY REMOVABLE
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues