Safemode Startup caused by a defective Z axis fan tray.
Symptom
Unable to Scan with Safemode Startup of XGD Failed for each Axis See Long Description Unable to scan with errors logged "Safemode Startup of XGD Failed" for every gradient axis seen at same time in error log. See errors, below. Notice that the date/time is the same for each error, and that all axes are reporting the error: X, Y & Z. Fri Mar 27 11:31:06 2009 Error: 2267798 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267801 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267804 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267807 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267810 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267813 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .->
Error message
2267834 The Power Cabinet AUX board is reporting a fan tachometer failure on # axis See Long Description Error: 2267834 Host: B4 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: mcgrafidy Coil: MCGRAFIDY1-6 Scan: 52532/1 The Power Cabinet AUX board is reporting a fan tachometer failure on # axis. Where # is X, Y, or Z axis.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5183212 FAN TRAY REMOVABLE
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.