Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact.

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2267827 XGD hardware warning detected. The PDU_X_INT_FAIL bit in AUX_SEC_FAULTS register was detected on the XGD GMC board. See Long Description Error: 2267827 Host: chmr2 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: depi_peddti Coil: 8HRBRAIN Scan: 4428/5 XGD hardware warning detected. The PDU_X_INT_FAIL bit in AUX_SEC_FAULTS register was detected on the XGD GMC board.

2267885 The % axis XPS is reporting a PLL Warning See Long Description Error: 2267885 Host: signa2 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The % axis XPS is reporting a PLL Warning.-> Extended Error: Ensure the fiber optic cable J4 is connected to the corresponding XPS, and that light is present. Check that the corresponding channel on the XPS-CB in the Cronus chassis has a good fiber optic output.

2267827 XGD hardware warning detected. The %s bit in %s register was detected on the %b %b. See Long Description Error: 2267827 Host: SLMAINMR Process: NSPFile: SCP:FspMac.cpp Line: 320Exception Class: Unknown Severity: UnknownFunction: Gradi

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.