IRF3 is Not Ready Due to PDU 48V power supply.
Symptom
Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.
Error message
2267446 The XGD module did not send a Fiber Link Init Complete 2267431 The IRF3 is not ready See Long Description Error: 2267431 Host: MR02 Process: NSPFile: AGP:Irf3RrfHWO Line: 1209Exception Class: Unknown Severity: UnknownFunction: Software/Firmware/Config: N/APSD: Unknown Coil: Unknown Scan: -1/-1 The IRF3 is not ready.-> Extended Error: This is just a warning message that indicates that another error message has already been logged about a serious problem with the IRF3, DTX, RRX, TDM, interconnect cable, or other error that is preventing the system from controlling the exciters and receivers. Look for a previous error message about configuration errors, cable errors, module errors, or calibration errors. Scans will not be allowed until that problem is corrected 2267414 NO XGD MODULE FOUND. The IRF3 board does not see the XGD module. See Long Description Error: 2267414 Host: PHSRADMR02 Process: diagCoordFile: Irf3.cpp Line: 2233Task ID: 0x493f570 Time: 25411 Object: AGPException Class: NotificationFunction: Software/Firmware/Config: N/A NO XGD MODULE FOUND.The IRF3 board does not see the XGD module. The system must have a dual fiber cable between the GMC board (in the Cronus
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.