GE Healthcare · MROptima2254207
Will not Scan with 2254207 SRFD2 RF Amplifier Device Fault 44 RF FET device is shorted Due to the SRFD2 Amplifier.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2254207 SRFD2 RF Amplifier Device Fault 44. See Long Description 2254207 RF FET device is shorted. SRFD2 will attempt to bypass this failure and run in low power mode. Replace Amplifier.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required