GE Healthcare · MROptimaDiscovery2268900
Unable to Scan with RF Air Temperature Faults Due To Customer Coolant Chiller.
Symptom
Unable to Scan with RF Air Temperature Faults
Error message
2268900 RF airflow warning detected. Future scanning will be inhibited until airflow matches expected. See Long Description 2268900 NSP SCP:SriCanHWO 2030 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 RF airflow warning detected. Future scanning will be inhibited until airflow matches expected. Airflow difference: 93 fpm Tolerance: 50 fpm-> Extended Error: The Heat Exchanger (HEC) manages closed loop control of the RF airflow this message indicates that the HEC was unable to adjust the frequency of the blower to achieve the requested airflow in the magnet.
Applies to
- Models
- Optima, Discovery
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.