2254781 Invalid power fault due to failed rf detector 2 board in UPM.

Symptom

Unable to scan with Power Monitor Faults

Error message

2254781 The Universal Power Monitor 1 has detected an Invalid Power Fault on the Narrow Band Body Amplifier. Wed Jul 18 10:06:06 2007 Error: 2254781 Host: MR3T Process: NSP File: SCP:UPM Line: 604 The Universal Power Monitor 1 has detected an Invalid Power Fault on the Narrow Band Body Amplifier.->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5250154 RF detector 2 board 5250154 RF detector 2 board
Area
MR