GE Healthcare · MROptima2254781
2254781 Invalid power fault due to failed rf detector 2 board in UPM.
Symptom
Unable to scan with Power Monitor Faults
Error message
2254781 The Universal Power Monitor 1 has detected an Invalid Power Fault on the Narrow Band Body Amplifier. Wed Jul 18 10:06:06 2007 Error: 2254781 Host: MR3T Process: NSP File: SCP:UPM Line: 604 The Universal Power Monitor 1 has detected an Invalid Power Fault on the Narrow Band Body Amplifier.->
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5250154 RF detector 2 board 5250154 RF detector 2 board
- Area
- MR