2267502 DTX calibration was aborted because the measured full strength signal was too low due to RFSB board in RF HUB.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2267502 Low signal during DTX1 calibration See Long Description 2267502 ERROR: EM_IRF3_DTX_CAL_LOW_SIGNAL Low signal during DTX%d calibration.DTX calibration was aborted because the measured full strength signal was too low. Expected signal level: %d, Actual level: %d Extended Error: With the DTX set to full gain, the measured signal level was too low.This signal is sent from the DTX to the RF Hub which loops it back to all the RRX input channels. A low signal can be caused by a bad DTX, a bad or disconnected cable, a defect in the RF Hub, or a bad RRX.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
6250052 RF Hub Switch Board RFSB 1 board required for 16 channel system, 2 boards required for 32 channel systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.