Gradient Coil coolant pump is leaking coolant.Unable to Scan with Gradient Coolant Flow and Level too Low.Replace Pump 5…
Gradient Coil coolant pump is leaking coolant.Unable to Scan with Gradient Coolant Flow and Level too Low.Replace Pump 5HP Pump Kit 5156321 wqr.Optima MR450W.2261712 Heat Exchanger Cabinet HEC detecte…
Symptom
Unable to Scan with Gradient Coolant Flow and Level too Low
Error message
2261712 Heat Exchanger Cabinet HEC detected the gradient coil coolant flow is too low Fri Apr 3 12:50:19 2009 Error: 2261712 Host: RIC_HDMR2 Process: NSP File: SCP:CoolingHWO Line: 2836 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: cal Coil: BODY Scan: 50270/1 Heat Exchanger Cabinet (HEC) detected the gradient coil coolant flow is too low. Verify that the pump is turned on, all in-line valves are open. This alarm will stop scanning 2261717 Heat Exchanger Cabinet (HEC) detected the coolant in the gradient coil reservoir is at the alarm level. 2261717 Host: chmr2 Process: NSP File: SCP:CoolingHWO Line: 2942 Exception Class: Unknown Severity: Unknown Function: Power/Support Systems: Heat Exchanger PSD: Unknown Coil: Unknown Scan: -1/-1 Heat Exchanger Cabinet (HEC) detected the coolant in the gradient coil reservoir is at the alarm level. Possible catastrophic coolant loss. Pumps are turned off automatically to protect the hardware. Please contact service. This alarm will stop scanning-> Extended Error: Please verify that there are no leaks in the system. If the pump isn''''''''t off, then check the level of coolant in the reservoir. If level
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5174313-4 FRU 4 gallons coolant non-copper 5156321 5HP PUMP KIT 5156321 5HP PUMP KIT
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.