GE Healthcare · CTDiscovery260118362956559655956559630956507840956466275956515840
Loud noise as system prepares to make xray, or start or brake the rotor.
Symptom
Loud noise as system prepares to make xray, or start or brake the rotor. Loud noise as system prepares to make xray for a tube warmup, scout or helical scan
Error message
Error: 260118362 Rotor Event Timestamps See Long Description for details Mon Jan 22 15:11:07 2007 Error: 260118362 Host: orp Process: orp_subsys File: gen_exec.c Line: 484 Mon Jan 22 15:11:06 2007 Host : On-Board Rotating Processor Ermes # : 260118362 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.c Line# : 484 Function : No System Function Reported Scan Type : Axial Scan: 4987/2/1 Rotor Event Timestamps(mS) Cur Time: 956559655 Rx: 956559630 Active Scan: 0 ExpDly End: 956559630 Accel: 956507840 Run: 956466275 Accel Time: 6000 Decel: 956515840 Stop: 956516025 Decel Time: 185 No More Rx: 956555175 End Exam: 956146975 Error: 260118324 Unable to bring up rotor due to acceleration interval constraint. (This is a message, not an error - See Long Description) See the Long Description for full error code When you get this error look for previous errors in the log. This error is usually the effect of another error and not the root cause. (example: if a tube over temperature is the root cause, you will get this error if another scan is immediately attempted after the over temperature error. The cooling delay allows the rotor to cool between acceleration attempts, but is not
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3701T : PERSEUS TUBE UNIT - Used on CT750 HD See Long Description for details. Part #: D3701T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERSEUS TUBE UNIT (GREEN) FRU: Yes Status: Orderable Comments: Per HDVCT RSPL, 212008898, 8/19/08. ** FE ASSIST ** Important to have adequate HV oil in the HV Wells. Making sure some oil does overflow when inserting Candle stick. (212001261 4/6/09)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.